Description
Intrinsically Safe Temperature Data Logger, Fixed RTD, -50 To 160°C, 3/16×12 with Data Logging
The best substitute for precision mercury- or spirit-filled glass thermometers. The RTD probe and thermometer readout are combined into one unit and calibrated as a system. Intrinsically safe and accurate to ±0.05 °C over full range. The 1551A features a robust thin-film RTD sensor, ranges from -50 °C to 160 °C and primarily targets the petroleum industry. Of course many other industries have applications that require an easy-to-use, stable and portable reference thermometer and the “Stik” Thermometer is the perfect solution.
Features:
-50 to 160 °C
Accurate to ± 0.05 °C (±0.09 °F) over full range
Intrinsically safe (ATEX and IECEx compliant)
Display temperature in °C or °F
User-configurable temperature Trend/Stability indicator
Optional data logging to internal memory available
300-hour battery life
Percent battery-life and low-battery indicator
NVLAP-accredited, NIST-traceable calibration includedIntrinsically Safe Temperature Data Logger, Fixed RTD, -50 To 160°C, 3/16×12 with Data Logging
The best substitute for precision mercury- or spirit-filled glass thermometers. The RTD probe and thermometer readout are combined into one unit and calibrated as a system. Intrinsically safe and accurate to ±0.05 °C over full range. The 1551A features a robust thin-film RTD sensor, ranges from -50 °C to 160 °C and primarily targets the petroleum industry. Of course many other industries have applications that require an easy-to-use, stable and portable reference thermometer and the “Stik” Thermometer is the perfect solution.
Features:
-50 to 160 °C
Accurate to ± 0.05 °C (±0.09 °F) over full range
Intrinsically safe (ATEX and IECEx compliant)
Display temperature in °C or °F
User-configurable temperature Trend/Stability indicator
Optional data logging to internal memory available
300-hour battery life
Percent battery-life and low-battery indicator
NVLAP-accredited, NIST-traceable calibration included
Specifications:
Specification Fluke 1551A Ex
Temperature range
-50 °C to 160 °C
(-58 °F to 320 °F)
Accuracy (1 year)
± 0.05 °C (± 0.09 °F)
Display units
°C, °F
Sensor type
100 Ω thin-film RTD
Probe temperature coefficient
0.00385 Ω/Ω/°C nominal
Sensor length
< 10 mm
Sensor position (from sheath tip)
3 mm (0.1 in)
Minimum immersion depth1
7 cm (2.8 in)
Probe sheath material
SST
Response time
4.8 mm diameter probe (3/16 in): 14 seconds
6.35 mm diameter probe (1/4 in): 21 seconds
Probe hysteresis
± 0.01 °C
Temperature resolution
Selectable: 0.1, 0.01, 0.001 (default 0.01
Sample rate
Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds
Operating temperature range readout
-10 °C to 50 °C (14 °F to 122 °F)
Storage temperature
-20 °C to 60 °C (-4 °F to 140 °F)
0 % to 95 % RH, non-condensing
Optional data logging2
Up to 10,000 time-stamped readings stored to internal memory
Logging intervals2
2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds; 2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes
Damping
Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10)
Communications
RS-232 stereo jack (access calibration parameters only)
AC power
None
DC power
Three AAA batteries, typical battery life of 300 hours without LCD backlight
EMC compliance
EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class "B"
Enclosure rating
IP50
Electronics dimensions
(H x W x D)
114 mm x 57 mm x 25 mm (4.5 in x 2.25 in x 1.0 in)
Weight
196 g (6.9 oz)
Calibration (included)
NVLAP-accredited, NIST-traceable
Characterization
CVD
1 Per ASTM E 644
2 See ordering information for optional data logging configurations
Click here for complete specifications on the Fluke 1551A-9-DL
What's included with the Fluke 1551A-9-DL
Thermometer
NVLAP-Accredited Report of Calibration
3 x AAA Batteries
User's Guide on CD-ROM
Reviews