Description
Fluke 725US-NIST OffersThis product includes an ISO Certificate with full Data and Uncertainties.
Features
Powerful, easy-to-use field calibrator
Measures volts, mA, pressure, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
Sources/simulates volts dc, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
Measures/sources pressure using any of 50 750P Pressure Modules (sold separately)
Easy to read measure/source screen displays input and output simultaneously
Powers transmitters during test using loop supply with simultaneous mA measurement
Handles fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
Provides a backlit display to provide better visibility in poor light
Features small, streamlined shape, and rugged, reliable design that stands up to field use
Sources mA with simultaneous pressure measurement to conduct valve and I/P tests
Supports flow meter testing with frequency and CPM functions
Performs fast linearity tests with auto-step and auto-ramp featuresFluke 725US-NIST OffersThis product includes an ISO Certificate with full Data and Uncertainties.
Features
Powerful, easy-to-use field calibrator
Measures volts, mA, pressure, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
Sources/simulates volts dc, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
Measures/sources pressure using any of 50 750P Pressure Modules (sold separately)
Easy to read measure/source screen displays input and output simultaneously
Powers transmitters during test using loop supply with simultaneous mA measurement
Handles fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
Provides a backlit display to provide better visibility in poor light
Features small, streamlined shape, and rugged, reliable design that stands up to field use
Sources mA with simultaneous pressure measurement to conduct valve and I/P tests
Supports flow meter testing with frequency and CPM functions
Performs fast linearity tests with auto-step and auto-ramp features
Fluke 725US-NIST Specifications
Measurement Accuracy
Voltage DC
30.000 V: 0.02% + 2 counts (upper display)
30.000 V: 0.02% + 2 counts (lower display)
100.00 mV: 0.02% + 2 counts
-10.00 to 75.00 mV: 0.025% + 1 count (via TC connector)
Current DC
24.000 mA: 0.02% + 2 counts
Resistance
0.0 to 400.0 Ω: 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω: 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω: 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency
2.0 to 1000.0 CPM: 0.05% + 1 count
1.0 to 1100.0 Hz: 0.05% + 1 count
1.00 to 10.00 kHz: 0.05% + 1 count
Sensitivity: 1 V peak-to-peak-minimum
Pressure
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC
100.00 mV: 0.02% + 2 counts
10.000 V: 0.02% + 2 counts
-10.00 to 75.00 mV: 0.025% + 1 count (via TC connector)
Current DC
24.000 mA (source): 0.02% + 2 counts
24.000 mA (simulate): 0.02% + 2 counts
Resistance
15.0 to 400.0 Ω: 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω: 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω: 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency
2.0 to 1000.0 CPM: 0.05%
1.0 to 1100.0 Hz: 0.05%
1.00 to 10.00 kHz: 0.25%
Waveform: 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure Accuracy
NI-120: 32.36°F (0.2°C)
PT-100 (385): 32.6°F (0.33°C)
PT-100 (393): 32.54°F (0.3°C)
PT-100 (JIS): 32.54°F (0.3°C)
PT-200 (385): 32.36°F (0.2°C)
PT-500 (385): 32.54°F (0.3°C)
PT-1000 (385): 32.36°F (0.2°C)
Resolution: 0.1°C
J: 33.26°F (0.7°C)
K: 33.44°F (0.8°C)
T: 33.44°F (0.8°C)
E: 33.26°F (0.7°C)
R: 35.24°F (1.8°C)
S: 34.7°F (1.5°C)
B: 34.52°F (1.4°C)
L: 33.26°F (0.7°C)
U: 33.35°F (0.75°C)
N: 33.62°F (0.9°C)
Resolution: J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK: 33.08°F (0.6°C)
BP: 34.16°F (1.2°C)
Source Accuracy
NI-120: 32.36°F (0.2°C)
PT-100 (385): 32.6°F (0.33°C)
PT-100 (393): 32.54°F (0.3°C)
PT-100 (JIS): 32.54°F (0.3°C)
PT-200 (385): 32.36°F (0.2°C)
PT-500 (385): 32.54°F (0.3°C)
PT-1000 (385): 32.36°F (0.2°C)
Resolution: 0.1°C
Accuracy stated for 4-wire measurement
J: 33.26°F (0.7°C)
K: 33.44°F (0.8°C)
T: 33.44°F (0.8°C)
E: 33.26°F (0.7°C)
R: 34.52°F (1.4°C)
S: 34.7°F (1.5°C)
B: 34.52°F (1.4°C)
L: 33.26°F (0.7°C)
U: 33.35°F (0.75°C)
N: 33.62°F (0.9°C)
Resolution: J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK: 33.08°F (0.6°C)
BP: 34.16°F (1.2°C)
Functions Specifications
Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature
Ramps: Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage: 24 V
Accuracy: 10%
Maximum current: 22 mA, short circuit protected
Step Functions
Source functions: Voltage, current, resistance, frequency, temperature
Steps: 25% of range, 100% of range
Environmental Specifications
Operating Temperature
14 to 131°F (-10 to 55°C)
Storage Temperature
-4 to 159.8°F (-20 to 71°C)
Operating Altitude
9842.52′ (3000 m)
Mechanical and General Specifications
Batteries
4 x AA alkaline batteries
Battery Replacement
Separate battery compartment, accessable without breaking calibration seal
Side Port Connections
Pressure module connector, also used for remote real-time programming
Dimensions
5.1 x 9.3 x 2.4″ (130 x 236 x 61 mm)
Weight
1.43 lbs (0.65 kg)
Safety Specifications
Agency Approvals
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Click here for complete specifications on the Fluke 725US-NIST
What’s included with the Fluke 725US-NIST
Process Calibrator
ISO Certificate with full Data and Uncertainties
Test Leads
Test Clips
Stackable Test Leads
Users Manuals
Statement of Quality Assurance Practices
CE and CSA Markings
Webinar: Pressure Calibrations with Fluke
Pressure devices provide critical measurement information to help your production control systems manage product quality, production outputs and activate critical safety systems. Keeping these pressure instruments operating within expected limits requires periodic verification, maintenance, and calibration.
This valuable one-hour Fluke webinar will give you the knowledge to quickly identify improvements to your pressure calibrations and associated workflows. Fluke provides you with key insights, the latest “best practice” methods and overview of the tools that will allow you to maximize results when testing the most common pressure instruments.
During this 1-hour webinar, we’ve covered:
•Pressure calibration – an overview and review
•Practical considerations when selecting calibration gear
•Best practices and tech tips
•Using technology for workflow improvements
•Maximize your efforts and achieving better results
Webinar: Temperature Calibrations with Fluke
We understand the importance of the performance of temperature and pressure instruments in your business as often being critical when trying to optimize operations, product quality or to ensure proper functioning of safety systems. As such, we’ve teamed up with Fluke to offer you a series of two webinars on this topic. The first will provide valuable information to help you identify temperature calibration improvements in your business. The second will focus on identifying pressure calibration improvements.
Let’s start with Temperature Calibrations… This valuable one-hour Fluke seminar will give you the knowledge to quickly identify improvements to your temperature calibrations and associated workflows. Providing you with key insights when reviewing your temperature calibration workload and highlights the latest “best practices”, methods and tools that will allow you to maximize results when testing the most common temperature instruments.
During this 1-hour webinar, we’ve covered:
• Temperature calibration – an overview and review
• Workload and applications checklist
• Identify, isolate and eliminate common sources of error
• Best practices and tech tips
• Using technology for workflow improvements
• How to maximize your results
Reviews